All studies represented by the chromatogram below have been conducted utilizing Vacuum Assisted Sorbent Extraction & Headspace Sorbent PensTM

The most versatile extraction and sample preconcentration technology available for GC and GCMS systems. Sorbent PensTM combine the features of SPME and classical ¼” adsorbent traps in a design with far more flexibility and enhanced performance. The Sorbent PenTM uses an entirely new approach for sample extraction.

Extremely Clean Blank After Strawberry analysis by VASE & Headspace Sorbent Pens

h

Instrument: 5800 HSP
Run date: August 3, 2016
Sample description: Fresh strawberries
Weight of sample (g): 7.7425
Sample conditions: 5 hours equilibration
Desorb conditions: 260°C for 5 min.
Split Mode: Splitless
Precolumn: DB1 5m length x 0.25mm ID, 0.25µm film
Column: DB1 30m length x 0.25mm ID, 0.5µm film
Carrier: He, 1.5cc/min. constant flow
Oven Temp: 40°C hold 5min., 6°C/min. to 95°C, 10°C/min. to 140°C, 15°C/min. to 325 hold 5 min.
GCMS: Agilent 7890B/5977A
MS Operation: 34-450 amu, 1.8 scans/sec

1. Ethyl acetate
2. Butanoic acid methyl ester
3. Propanoic acid, 2-methyl-, ethyl ester
4. Butanoic acid, ethyl ester
5. Acetic acid, butyl ester
6. Butanoic acid, 2-methyl-, ethyl ester
7. 1-Butanol, 3-methyl-, acetate
8. Hexanoic acid, methyl ester
9. Hexanoic acid, ethyl ester
10. 2-Hexen-1-ol, acetate, (Z)-

11. 3(2H)-Furanone, 4-methoxy-2,5-dimethyl-
12. 1,6-Octadien-3-ol, 3,7-dimethyl-
13. Octanoic acid, methyl ester
14. Methyl salicylate
15. Octanoic acid, ethyl ester
16. Butanoic acid, octyl ester
17. Pentanoic acid, octyl ester
18. 1,6,10-Dodecatrien-3-ol, 3,7,11-trimethyl-
19. .gamma.-Dodecalactone