All studies represented by the chromatogram below have been conducted utilizing Vacuum Assisted Sorbent Extraction & Headspace Sorbent PensTM

The most versatile extraction and sample preconcentration technology available for GC and GCMS systems. Sorbent PensTM combine the features of SPME and classical ¼” adsorbent traps in a design with far more flexibility and enhanced performance. The Sorbent PenTM uses an entirely new approach for sample extraction.

Mushroom analysis by VASE & Headspace Sorbent Pens

h

Instrument: 5800 HSP
Run date: August 22, 2016
Sample description: Mushrooms blended with water (1:2 ratio)Weight of Sample (mL): 7.7366
Weight of Sample (g): 2
Sample Conditions: 40 mL vial with 15 hours equilibration at 25°C
Desorb conditions: 260°C for 5 min.
Split Mode: Splitless
Precolumn: DB1 5m length x 0.25mm ID, 0.25µm film
Column: DB1 30m length x 0.25mm ID, 0.5µm film
Carrier: He, 1.5cc/min. constant flow
Oven Temp: 40°C hold 5 min., 10°C/min. to 325C, 325 hold 5 min. (solvent delay at 6 min)
GCMS: Agilent 7890B/5977A
MS Operation: 34-450 amu, 1.8 scans/sec

1. Acetic acid
2. 1-Butanol, 3-methyl-
3. Hexanal
4. 1-Hexanol
5. Benzaldehyde
6. 1-Octen-3-one
7. 3-Octanone
8. Octanal
9. Benzeneacetaldehyde
10. 2-Octenal, (E)-

11. 2-Octen-1-ol, (Z)-
12. Propanoic acid, 2-methyl-, 2,2-dimethyl-1- (2-hydroxy-1-methylethyl)propyl ester
13. 5,9-Undecadien-2-one, 6,10-dimethyl-, (E)-
14. Phenol, 2,4-bis(1,1-dimethylethyl)-
15. Benzophenone
16. 1,2-Benzenedicarboxylic acid, bis(2-methylpropyl) ester
17. Dibutyl phthalate
18. Isopropyl palmitate
19. Octadecanoic acid

The Entech web store is currently undergoing maintenance. Accounts may be temporarily disabled as well as access to quotes, account history, and support documents. Please email support@entechinst.com with any support documentation needs.