CS1200E Inlet Flow Restrictor Elements (Silonite Coated)

Available Options:

Product Description

Used in the Entech CS1200E(S) and other products.
Inlet restrictors provide stable operation from -20 ̊C to +50 ̊C. Now with 9 restrictor ranges covering 15 minutes to 1 month.

Additional Information

Weight 1 lbs
Flow Code

2+, 3+, O, 1, 2, 3, 5, 4+, 6, 4, 7

Flow Range

0.1-0.3 cc/min, 0.2-0.6 cc/min, 0.5-1.5 cc/min, 1-3 cc/min, 12-36 cc/min, 150-450 cc/min, 2-6 cc/min, 20-60 cc/min, 4-12 cc/min, 50-150 cc/min, 7-21 cc/min

Unit

EA

cs1200e

Ristrictor & Canister Fill Duration Range

CS1200 with stamped flow range code.

  Recommended Restrictor for Volumes and Fill Times
Fill Duration 450ML 600ML 1L 1.4L 2L 2.7L 3.2L 6L 15L
15 min 2+ 2 1 1 1 0 0 0
1 hr 3+ 3+ 3 2+ 2+ 2 2 1 0
3 hrs 4+ 4 4 3+ 3 3 3 2+ 1
8 hrs 5 5 4+ 4+ 4 4 3+ 3 2+
12 hrs 6 5 5 4+ 4+ 4 3+ 3+ 2+
1 day 6 6 5 5 5 4+ 4+ 4 3+
2 days 7 7 6 6 5 5 5 4+ 4
7 days 7 7 7 7 6 5
14 days 7 7 7 5
30 days 7 6
Assuming canisters are filled to 4″ Hg below atmosphere

Canister Target Flow rate and Fill Duration

Target Flow Rate show in cc/min. for canisters with Micro-QT valves and the Toxic Organics Valve only.

Fill Duration 450ML
cc/min
600ML
cc/min
1L
cc/min
1.4L
cc/min
2L
cc/min
2.7L
cc/min
3.2L
cc/min
6L
cc/min
15L
cc/min
1 Hour 6.5 8.7 14.5 20.3 39.2 46.4 87
2 Hours 3.3 4.4 7.3 10 20 23.2 43.5 109
3 Hours 2.2 2.9 4.8 6.8 13.1 15.5 29 72.5
4 Hours 1.63 2.2 3.6 5.1 9.8 11.6 21.8 54.4
8 Hours 0.82 1.1 1.8 2.5 4.9 5.8 10.9 27.2
12 Hours 0.54 0.73 1.2 1.7 3.3 3.9 7.3 18.2
24 Hours 0.60 0.85 1.6 1.9 3.6 9.06
2 Days 0.82 0.97 1.8 4.5
1 Week 0.52 1.3
For sampling events above 3000 feet, reduce flow rate by 15%. MicrovalvesTM are recommended for use on SiloniteTM Canisters and MinicansTM upto 3.2L. Larger Canisters with MicroValvesTM may require longer clearing time due to the added restriction of the valve. the Toxic Organics ValveTM can be used with all canister sizes with the exception of 600mL.

The 600mL MinicanTM with an integrated Micro-QT2TM valve is designed only for MicroValveTM connections.

The following tables shows the target flow rates of various canister sizes and sampling durations.
(Target pressure is 0.87 atmospheres)

450mL Minican – Time from 20″Hg to 10″Hg

Gauge Seconds Code Target Flow Rate
cc/min
Duration
15.5 4 2.21 3 Hours
31 5 1.09 6 Hours
41.5 5 0.82 8 Hours

1L Minican – Time from 20″Hg to 10″Hg

Gauge Seconds Code Target Flow Rate
cc/min
Duration
7 4 4.8 3 Hours
18.5 4 1.80 8 Hours
56 5 0.60 24 Hours

3.2L SiloniteTM Canister – Time from 20″Hg to 10″Hg

Gauge Seconds Code Target Flow Rate
cc/min
Duration
6 4 5.8 8 Hours
17.5 4 1.93 24 Hours
35 5 0.96 82 Days

6L SiloniteTM Canister – Time from 20″Hg to 10″Hg

Gauge Seconds Code Target Flow Rate
cc/min
Duration
5 3 7.3 12 Hours
9.5 4 3.6 24 Hours
65.5 5 0.52 1 week

CS1200

Entech CS1200 Flow Controller for Time integrated Sampling

Related Products