Silonite™ CS1200E Passive Canister Sampler

SKU: 39-CS1200ES

Now featuring all new low flow stability for extremely precise Time Weighted Average

Low Level TO-15 monitoring requires consistent analyte recovery while collecting the maximum sample volume possible to support larger preconcentration volumes or repeat analyses from the same canister. The CS1200E will reliably fill canisters at a constant rate with better demonstrated recovery of TO-15 compounds than any other sampler.

Login or register to see product details

Available Options:

Product Description

Now featuring all new low flow stability for extremely precise Time Weighted Average

Low Level TO-15 monitoring requires consistent analyte recovery while collecting the maximum sample volume possible to support larger preconcentration volumes or repeat analyses from the same canister. The CS1200E will reliably fill canisters at a constant rate with better demonstrated recovery of TO-15 compounds than any other sampler.

Time-weighted average VOC concentrations can be easily determined by sampling into canisters at a constant flow rate. Canisters can be filled to nearly atmospheric pressure using an ultra-clean, mechanical flow controller. Sapphire orifices combined with pressure regulation offer superior flow stability when compared to needle valve or frit regulated controllers. Different orifices are available to fill a 6L canister over 1, 3, 8 or 24 hours, or even 1–2 weeks. Part numbers for CS1200E flow controllers are separated by flow range. One flow range can be changed to another by attaching the appropriate critical orifice shown in the table below. Silonite® coated flow controllers are designated with an “S” following the part number.

Additional Information

Weight 1 lbs
Flow Range

0.2-0.6 cc/min, 0.1-0.3 cc/min, 150-350 cc/min, 20-60 cc/min, 8-20 cc/min, 4-9 cc/min, 2-6 cc/min, 1-3 cc/min, 0.35-1 cc/min, 50-150 cc/min, 150-450 cc/min, 7-21 cc/min, 4-12 cc/min, 0.5-1.5 cc/min

Ristrictor & Canister Fill Duration Range

CS1200 with stamped flow range code.

CS1200E | Restrictors Recommended Restrictor for Volumes and Fill Times
Code Part No. Flow Range Fill Duration 450ML 600ML 1L 1.4L 2.7L 3.2L 6L 15L
0 39-23000(S) 150-450cc/min 15 min 2 2 1 1 0 0 0
1 39-23010(S) 50-150cc/min 1 hr 3+ 3+ 3 3 2 2 1 0
2 39-23030(S) 20 – 60cc/min 3 hrs 4+ 4 4 3+ 3 3 2 1
3 39-23080(S) 7 – 21cc/min 8 hrs 5 5 4+ 4+ 4 3+ 3 2
3+ 39-23160(S) 4 – 12cc/min 12 hrs 6 5 5 4+ 4 4 3+ 3
4 39-23240(S) 2 – 6cc/min 1 day 6 6 5 5 4+ 4+ 4 3+
4+ 39-23480(S) 1 – 3cc/min 2 days 7 7 6 6 5 5 4+ 4
5 39-24010(S) 0.5-1.5cc/min 7 days 7 7 7 6 5
6 39-24020S* 0.2-0.6cc/min 14 days 7 7 7 5
7 39-24040S* 0.1-0.3cc/min 30 days 7 6

*Available in Silonite Coated version only.

Assuming canisters are filled to 4″ Hg below atmosphere

Canister Target Flow rate and Fill Duration

Target Flow Rate show in cc/min. for canisters with Micro-QT valves and the Toxic Organics Valve only.

Fill Duration 450ML
cc/min
600ML
cc/min
1L
cc/min
1.4L
cc/min
2.7L
cc/min
3.2L
cc/min
6L
cc/min
15L
cc/min
1 Hour 6.5 8.7 14.5 20.3 39.2 46.4 87
2 Hours 3.3 4.4 7.3 10 20 23.2 43.5 109
3 Hours 2.2 2.9 4.8 6.8 13.1 15.5 29 72.5
4 Hours 1.63 2.2 3.6 5.1 9.8 11.6 21.8 54.4
8 Hours 0.82 1.1 1.8 2.5 4.9 5.8 10.9 27.2
12 Hours 0.54 0.73 1.2 1.7 3.3 3.9 7.3 18.2
24 Hours 0.60 0.85 1.6 1.9 3.6 9.06
2 Days 0.82 0.97 1.8 4.5
1 Week 0.52 1.3

For sampling events above 3000 feet, reduce flow rate by 15%. MicrovalvesTM are recommended for use on SiloniteTM Canisters and MinicansTM upto 3.2L. Larger Canisters with MicroValvesTM may require longer clearing time due to the added restriction of the valve. the Toxic Organics ValveTM can be used with all canister sizes with the exception of 600mL.

The 600mL MinicanTM with an integrated Micro-QT2TM valve is designed only for MicroValveTM connections.

The following tables shows the target flow rates of various canister sizes and sampling durations.
(Target pressure is 0.87 atmospheres)

450mL Minican – Time from 20″Hg to 10″Hg

Gauge Seconds Code Target Flow Rate
cc/min
Duration
15.5 4 2.21 3 Hours
31 5 1.09 6 Hours
41.5 5 0.82 8 Hours

1L Minican – Time from 20″Hg to 10″Hg

Gauge Seconds Code Target Flow Rate
cc/min
Duration
7 4 4.8 3 Hours
18.5 4 1.80 8 Hours
56 5 0.60 24 Hours

3.2L SiloniteTM Canister – Time from 20″Hg to 10″Hg

Gauge Seconds Code Target Flow Rate
cc/min
Duration
6 4 5.8 8 Hours
17.5 4 1.93 24 Hours
35 5 0.96 82 Days

6L SiloniteTM Canister – Time from 20″Hg to 10″Hg

Gauge Seconds Code Target Flow Rate
cc/min
Duration
5 3 7.3 12 Hours
9.5 4 3.6 24 Hours
65.5 5 0.52 1 week