Tedlar® Bags are convenient for the analysis of
VOCs and light gases. Due to the semi-permeable
nature of Tedlar, analysis should be performed quickly
to avoid loses. Unattended sample injections occurring
over-night could make the difference in completing the
analysis during the acceptable holding period.
Automated GC sample injection can also improve reproducibility
by eliminating the possibility of sample loss due to
leaky injection port septa.
7032A-LB
Tedlar® Bag Autosampler
The 7032A-LB Autosampler automates the introduction
of up to 21 samples into a gas chromatograph for quantitative
analysis. A fixed volume loop is used to measure
and equilibrate sample volumes to atmospheric pressure
before injection into the GC. A vacuum pump draws
a sample from each bag while keeping the sample loop
at atmospheric pressure. The ability to maintain
consistent pressures, temperatures, and loop volumes
during the injection makes the 7032A-LB Autosampler
a highly reproducible analytical tool for analyzing
bag samples.
|
 |
ESS
is a premier supplier of Tedlar® Air Sampling Bags.
Order yours today! |
|
|
|
Variable Volume Injections
The 7032A-LB used an electronic splitter to adjust the
actual volume of the sample delivered to a GC.
Using a single loop, the volume can be split as much
as 100:1 allowing a greater range of sample concentrations
to be analyzed without dilution. Variable injection
volumes also allow calibration curves to be generated
from a single calibrated standard cylinder. Internal
standards can be added independent of samples or analytical
standards, providing added QA to the analysis.
|
|
Tool Free Sample Connection
The 7032A-LB uses black anodized torque wheels attached
to each fitting to allow finger tightening of the bags
during autosampler loading and unloading. This
improves laboratory efficiency and increases sample
throughput. Teflon reducing ferrules (1/4" to
3/16") provide a reliable seal and are heavy duty enough
to use over and over. Direct connections can be
made to both metal and plastic Tedlar® bag valves
as supplied by ESS and SKC.
|
|
Large Volume Preconcentration
The
7032A-LB is also available without a loop injection
valve (Model 7032A-B) for operation with the 7100A Preconcentrator.
Instead of being limited to a maximum of 5cc, the 7100A/7032A-B
Preconcentration System allows analysis of 10-1000cc.
This greatly enhances detection limits, allowing the
analysis of compounds in the 1-1000 ppb range.
|
|
Focusing option
In order to maintain narrow peak width and maximum sensitivity
when performing capillary GC analysis, it may be necessary
to focus the sample at the head of the column to eliminate
band broadening of the more volatile compounds.
The 7032-02 focusing option is capable of cooling to
-190†C using liquid nitrogen during each injection cycle.
Heating of the focuser upon injections can approach
10,000 deg/min, giving extremely fast release of the
sample to a capillary gas chromatograph.
|
|
System Back-flushing
The 7032A-LB utilizes fused silica lined tubing to reduce
transfer losses and retention of contaminants.
To further reduce carryover, each sample inlet can be
back-flushed with nitrogen for a user selected period
of time before loading the autosampler with the next
set of bags. This virtually eliminates the possibility
for system memory effects.
|
|
|
|
|
-
Landfill Gas
-
Remote Head Space
-
Mobile Laboratories
-
Petroleum/Petrochemical
|
-
Product Offgasing
-
Process Stream
-
Soil Gas
-
Indoor Air
|
|
|
|
|
Description
|
Part Number
|
|
21 Position Loop Injection Auto sampler
|
7032A-LB
|
|
21 Position Autosampler for Operation with 7100A
|
7032A-B
|
|
Electronic Splitter
|
7032-01
|
|
Electronic Splitter Flow Calibrator
|
39-20035
|
|
Focusing Option
|
7032-02
|
|
Option for 7032-LB to interface with 7100 Preconcentrator
|
7032-03
|
|