logo

The 3106A Automated Canister Cleaner

Fused silica lined canisters are the preferred means of collecting VOC’s at ppm to sub ppb levels in the gas phase. After analysis, the canisters must be cleaned for reuse by eliminating all gas-phase and surface bound contaminants.

Failure to remove all contaminants will cause carryover, false-positive detection and poor quantification. In addition, heavier organics remaining in the canisters can result in buildup which may interact with the sample.

 

* Canisters occasionally are contaminated beyond the capability of any vacuum canister cleaner. Click here to download our advanced cleaning procedure.

The 3106A canister cleaning system brings simplicity and effective quality control to the cleaning process. During cleaning, canister evacuation is performed in two stages. The first stage utilizes a diaphragm pump to perform a rough evacuation of the canisters. A molecular drag pump then further reduces the vacuum down to the millitorr range. Both pumps are oil free, eliminating the need for isolation traps. After evacuation, canisters are refilled with humidified zero air or nitrogen to dilute any remaining impurities. Cycling between fill and evacuation effectively eliminates all traces of VOC contaminants.

Up to 24 SUMMA® or Silonite® canisters can be cleaned at once. Canister connections are made to a dual-path Silonite® manifold. Separate fill and evacuation paths prevent backflushing of contaminates into the cleaned canisters. The inverted canister connections allow any size canister to be cleaned. Transducers provide digital indication of the rough pressure and vacuum (0-50 psia) as well as high vacuum (0-2000 mtorr).



SmartLabII™ Control Interface
The 3106A is controlled using the Entech SmartLab™ Control Network which is capable of controlling an entire air analysis laboratory from a single computer. Operating under Microsoft Windows™ 2000 or XP the interface includes screens for developing cleaning methods, running the system automatically or manually, and performing system diagnostics. Methods include selection of cleanup parameters, automatic leak checking, and required vacuum for final evacuation. During unattended operation, parameters are saved to a report for quality assurance and to maintain a permanent record of the cleaning and leak-checking event. As with the other SmartLab™ peripherals, system operation and vendor support is possible through remote modem communication.

Features

  • Oven sized to accommodates two-15 L canisters, six-6L or 3.2L canisters or
    eight-2.7L or smaller canisters.
  • Oven heating eliminates the need for multiple sizes of heating bands
  • Inverted cleaning position accommodates canisters from any vendor.
  • Control and pumping capacity for 4 ovens from a single 3100A module.
  • Separate fill and evacuation paths prevent backflushing of contaminates into the cleaned canisters.
  • Oilless ruffing and turbo pumps prevents pump oil contamination of canisters
  • Automated switching between ruffing and turbo pump, pressure and vacuum gauge
  • Computer reporting of system pressure (0-50psia) and vacuum (0-2000mtorr)
  • Automatic leak-checking during cleaning
  • SmartLabII™ USB/computer system control


 

Ordering Information
Description   Part Number
Model 3106 Canister Cleaner with 6 Position Canister Oven, Silonite® Dual-path Manifold and Entech SmartLab II Network 01-3106A
Upgrade Your Existing 3101 Cleaner to Dual-Path Oven Cleaner
3100 Upgrade Kit, SmartLabI to SmartLabII 90-83100

3100 2-stage manifold upgrade kit

(requires oven and manifold)

90-31002
Accessories - Expand Your 3106 to clean 24 SUMMA® style canisters

Multi-Oven Parallel Option

(one for each additional oven)

3100-03

6 Canister Oven (requires manifold)

Max 4 ovens per cleaner

09-OV6L
6 Canister Silonite® Dual-path Manifold (requires oven) 3100-06
Add MiniCan cleaning to your existing Entech 3106  
20-position MiniCan Oven (requires manifold) O9-OV20

20-position MiniCan Silonite® Dual-path Manifold

(requires oven)

3100-20