Fused
silica lined canisters are the preferred means of collecting
VOCs at ppm to sub ppb levels in the gas phase.
After analysis, the canisters must be cleaned for reuse
by eliminating all gas-phase and surface bound contaminants.
Failure
to remove all contaminants will cause carryover, false-positive
detection and poor quantification. In addition, heavier
organics remaining in the canisters can result in buildup
which may interact with the sample.
*
Canisters occasionally are contaminated beyond the capability
of any vacuum canister cleaner. Click here to download
our advanced cleaning procedure.
|
|
The
3106A canister cleaning system brings simplicity and
effective quality control to the cleaning process.
During cleaning, canister evacuation is performed
in two stages. The first stage utilizes a diaphragm
pump to perform a rough evacuation of the canisters.
A molecular drag pump then further reduces the vacuum
down to the millitorr range. Both pumps are oil free,
eliminating the need for isolation traps. After evacuation,
canisters are refilled with humidified zero air or
nitrogen to dilute any remaining impurities. Cycling
between fill and evacuation effectively eliminates
all traces of VOC contaminants.
Up
to 24 SUMMA® or Silonite® canisters can be
cleaned at once. Canister connections are made to
a dual-path Silonite® manifold. Separate fill
and evacuation paths prevent backflushing of contaminates
into the cleaned canisters. The inverted canister
connections allow any size canister to be cleaned.
Transducers provide digital indication of the rough
pressure and vacuum (0-50 psia) as well as high vacuum
(0-2000 mtorr).
|
|
SmartLabII Control Interface
The 3106A is controlled using the Entech SmartLab
Control Network which is capable of controlling an entire
air analysis laboratory from a single computer. Operating
under Microsoft Windows 2000 or XP the interface
includes screens for developing cleaning methods, running
the system automatically or manually, and performing
system diagnostics. Methods include selection of cleanup
parameters, automatic leak checking, and required vacuum
for final evacuation. During unattended operation, parameters
are saved to a report for quality assurance and to maintain
a permanent record of the cleaning and leak-checking
event. As with the other SmartLab peripherals,
system operation and vendor support is possible through
remote modem communication.
|
|
Features
- Oven
sized to accommodates two-15 L canisters, six-6L
or 3.2L canisters or
eight-2.7L or smaller canisters.
- Oven
heating eliminates the need for multiple sizes of
heating bands
- Inverted
cleaning position accommodates canisters from any
vendor.
- Control
and pumping capacity for 4 ovens from a single 3100A
module.
- Separate
fill and evacuation paths prevent backflushing of
contaminates into the cleaned canisters.
- Oilless
ruffing and turbo pumps prevents pump oil contamination
of canisters
- Automated
switching between ruffing and turbo pump, pressure
and vacuum gauge
- Computer
reporting of system pressure (0-50psia) and vacuum
(0-2000mtorr)
- Automatic
leak-checking during cleaning
- SmartLabII
USB/computer system control
|
|
Ordering
Information
|
| Description |
|
Part
Number |
| Model
3106 Canister Cleaner with 6 Position Canister Oven, Silonite®
Dual-path Manifold and Entech SmartLab II Network |
01-3106A |
| Upgrade
Your Existing 3101 Cleaner to Dual-Path Oven Cleaner |
| 3100
Upgrade Kit, SmartLabI to SmartLabII |
90-83100 |
|
3100
2-stage manifold upgrade kit
(requires
oven and manifold)
|
90-31002 |
| Accessories
- Expand Your 3106 to clean 24 SUMMA® style canisters |
|
Multi-Oven
Parallel Option
(one
for each additional oven)
|
3100-03 |
|
6
Canister Oven (requires manifold)
Max
4 ovens per cleaner
|
09-OV6L |
| 6
Canister Silonite® Dual-path Manifold (requires oven) |
3100-06 |
| Add
MiniCan cleaning to your existing Entech 3106 |
|
| 20-position
MiniCan Oven (requires manifold) |
O9-OV20 |
|
20-position
MiniCan Silonite® Dual-path Manifold
(requires
oven)
|
3100-20 |